Outcomes Beyond 10 Years After Transcatheter Aortic Valve Implantation in High-Risk Patients With Severe Aortic Valve Stenosis

Research output: Contribution to journal › Research article › Contributed › peer-review

Contributors

  • Ellen Dietze - , Heart Center Leipzig (Author)
  • Ines Richter - , Heart Center Leipzig (Author)
  • Oliver Dumpies - , Heart Center Leipzig (Author)
  • Johannes Rotta Detto Loria - , Heart Center Leipzig (Author)
  • Nicolas Majunke - , Heart Center Leipzig (Author)
  • Hans-Josef Feistritzer - , Heart Center Leipzig (Author)
  • Philipp Kiefer - , Heart Center Leipzig (Author)
  • David Holzhey - , Heart Center Leipzig (Author)
  • Thilo Noack - , Heart Center Leipzig (Author)
  • Axel Linke - , Department of Internal Medicine and Cardiology (at Dresden Heart Centre) (Author)
  • Norman Mangner - , Department of Internal Medicine and Cardiology (at Dresden Heart Centre) (Author)
  • Steffen Desch - , Heart Center Leipzig (Author)
  • Michael Borger - , Heart Center Leipzig (Author)
  • Holger Thiele - , Heart Center Leipzig (Author)
  • Mohamed Abdel-Wahab - , Heart Center Leipzig (Author)

Abstract

BACKGROUND: Limited data is available on long-term outcomes and valve durability measures of transcatheter aortic valve implantation (TAVI).

AIMS: This study sought to assess clinical and echocardiographic outcomes of high-risk patients during the early experience of TAVI with a follow-up period extending beyond 10 years.

METHODS: Patients were included who had undergone TAVI at the Heart Center Leipzig between 2006 and 2012. Valve durability measures including moderate and severe structural valve deterioration (SVD) were defined according to the updated standardized definitions of the Valve Academic Research Consortium 3 (VARC-3).

RESULTS: A total of 1825 patients were included. The 12-year mortality rate was 92.8%. At 12 years post-TAVI, the cumulative incidence of SVD was 9.8% in the total population. Complete echocardiographic follow-up at ≥ 9 years was available for 56 patients. The median follow-up of this subgroup was 10.0 years. Moderate SVD was observed in three patients of the echocardiographic follow-up cohort (5.4%) and severe SVD in two patients (3.6%). Self-expanding (SE) transcatheter heart valves (THV) had a significantly lower rate of any SVD (2.8% vs 20.0%, p = 0.030) than balloon-expandable (BE) THV.

CONCLUSION: Long-term mortality rates in high-risk patients treated with TAVI more than 10 years ago are substantial, which limits assessment of valve durability measures. In a subgroup of survivors beyond 9 years, the cumulative incidence of SVD was low.

Details

Original languageEnglish
Pages (from-to)1301-1309
Number of pages9
JournalCatheterization and Cardiovascular Interventions
Volume106
Issue number2
Publication statusPublished - 1 Aug 2025
Peer-reviewedYes

External IDs

PubMedCentral PMC12336767
Scopus 105008205185

Keywords

Keywords

  • Humans, Aortic Valve Stenosis/mortality, Male, Female, Time Factors, Treatment Outcome, Aged, 80 and over, Transcatheter Aortic Valve Replacement/adverse effects, Risk Factors, Severity of Illness Index, Aortic Valve/diagnostic imaging, Aged, Heart Valve Prosthesis, Germany/epidemiology, Risk Assessment, Prosthesis Design, Retrospective Studies, Prosthesis Failure