On the measurement of total radiated power in uncalibrated reverberation chambers
Publikation: Beitrag in Fachzeitschrift › Forschungsartikel › Beigetragen › Begutachtung
Beitragende
Abstract
A novel experimental procedure to measure the total radiated power of an equipment under test (EUT) placed in a reverberation chamber (RC) is presented. In contrast to the well-established method of the IEC 61000-4-21, this new procedure does not rely on information obtained during the empty chamber calibration or the calibration with the EUT in place. Thus, the method is simpler, faster, and may have a smaller uncertainty budget.
Details
Originalsprache | Englisch |
---|---|
Seiten (von - bis) | 270-279 |
Seitenumfang | 10 |
Fachzeitschrift | IEEE Transactions on Electromagnetic Compatibility |
Jahrgang | 49 |
Ausgabenummer | 2 |
Publikationsstatus | Veröffentlicht - Mai 2007 |
Peer-Review-Status | Ja |
Externe IDs
ORCID | /0000-0002-3894-9831/work/142252666 |
---|
Schlagworte
ASJC Scopus Sachgebiete
Schlagwörter
- Emission measurement, IEC 61000-4-21, Mode-stirred chamber, Reverberation chamber (RC), Total radiated power