Report from the HarmoSter study: different LC-MS/MS androstenedione, DHEAS and testosterone methods compare well; however, unifying calibration is a double-edged sword

Research output: Contribution to journalResearch articleContributedpeer-review

Contributors

  • Flaminia Fanelli - , University of Bologna (Author)
  • Mirko Peitzsch - , Institute of Clinical Chemistry and Laboratory Medicine (Author)
  • Stephen Bruce - , University Hospital of Lausanne (CHUV) (Author)
  • Marco Cantù - , Ente Ospedaliero Cantonale (EOC) (Author)
  • Anastasia Temchenko - , University of Bologna (Author)
  • Marco Mezzullo - , University of Bologna (Author)
  • Johanna M Lindner - , Hospital of the Ludwig-Maximilians-University (LMU) Munich (Author)
  • James M Hawley - , Manchester University NHS Foundation Trust (Author)
  • Mariette T Ackermans - , University of Amsterdam (Author)
  • Jody Van den Ouweland - , Canisius Wilhelmina Hospital (Author)
  • Daniel Koeppl - , University Hospital at the Friedrich-Alexander University Erlangen-Nürnberg (Author)
  • Elena Nardi - , IRCCS Azienda Ospedaliero-Universitaria di Bologna (Author)
  • Finlay MacKenzie - , University Hospitals Birmingham NHS Foundation Trust (Author)
  • Pierre-Alain Binz - , University Hospital of Lausanne (CHUV) (Author)
  • Manfred Rauh - , University Hospital at the Friedrich-Alexander University Erlangen-Nürnberg (Author)
  • Brian G Keevil - , Manchester University NHS Foundation Trust (Author)
  • Michael Vogeser - , Hospital of the Ludwig-Maximilians-University (LMU) Munich (Author)
  • Graeme Eisenhofer - , Institute of Clinical Chemistry and Laboratory Medicine (Author)
  • Annemieke C Heijboer - , Amsterdam University Medical Centers (UMC), University of Amsterdam, Vrije Universiteit Amsterdam (VU) (Author)
  • Uberto Pagotto - , IRCCS Azienda Ospedaliero-Universitaria di Bologna (Author)

Abstract

OBJECTIVES: Current liquid chromatography-tandem mass spectrometry (LC-MS/MS) applications for circulating androgen measurements are technically diverse. Previously, variable results have been reported for testosterone. Data are scarce for androstenedione and absent for dehydroepiandrosterone sulfate (DHEAS). We assessed the agreement of androstenedione, DHEAS and testosterone LC-MS/MS measurements among nine European centers and explored benefits of calibration system unification.

METHODS: Androgens were measured twice by laboratory-specific procedures in 78 patient samples and in EQA materials. Results were obtained by in-house and external calibration. Intra- and inter-laboratory performances were valued.

RESULTS: Intra-laboratory CVs ranged between 4.2-13.2 % for androstenedione, 1.6-10.8 % for DHEAS, and 4.3-8.7 % and 2.6-7.1 % for female and male testosterone, respectively. Bias and trueness in EQA materials were within ±20 %. Median inter-laboratory CV with in-house vs. external calibration were 12.0 vs. 9.6 % for androstenedione (p<0.001), 7.2 vs. 4.9 % for DHEAS (p<0.001), 6.4 vs. 7.6 % for female testosterone (p<0.001) and 6.8 and 7.4 % for male testosterone (p=0.111). Median bias vs. all laboratory median with in-house and external calibration were -13.3 to 20.5 % and -4.9 to 18.7 % for androstenedione, -10.9 to 4.8 % and -3.4 to 3.5 % for DHEAS, -2.7 to 6.5 % and -11.3 to 6.6 % for testosterone in females, and -7.0 to 8.5 % and -7.5 to 11.8 % for testosterone in males, respectively.

CONCLUSIONS: Methods showed high intra-laboratory precision but variable bias and trueness. Inter-laboratory agreement was remarkably good. Calibration system unification improved agreement in androstenedione and DHEAS, but not in testosterone measurements. Multiple components, such as commutability of calibrators and EQA materials and internal standard choices, likely contribute to inter-laboratory variability.

Details

Original languageEnglish
Pages (from-to)1080-1091
Number of pages12
JournalClinical chemistry and laboratory medicine
Volume62
Issue number6
Publication statusPublished - 27 May 2024
Peer-reviewedYes

External IDs

Scopus 85182563203

Keywords

Keywords

  • Androstenedione/blood, Calibration, Dehydroepiandrosterone Sulfate/blood, Female, Humans, Liquid Chromatography-Mass Spectrometry/methods, Male, Middle Aged, Tandem Mass Spectrometry/standards, Testosterone/blood