LithoVue Elite Compared With Single-Use and Reusable Ureteroscopes: A Bench and Human Cadaver Study

Research output: Contribution to journalResearch articleContributedpeer-review

Contributors

  • Naeem Bhojani - , Centre Hospitalier de l'Université de Montreal (CHUM) (Author)
  • Nabil Shalabi - , University of British Columbia (Author)
  • Alina Reicherz - , University of British Columbia (Author)
  • Roman Herout - , Department of Urology, University of British Columbia (Author)
  • Kyle Searles - , University of British Columbia (Author)
  • K. f. victor Wong - , University of British Columbia (Author)
  • Ben h. Chew - , University of British Columbia (Author)
  • Marien Hospital Herne

Abstract

Purpose:
To compare the attributes of the next-generation LithoVue Elite (LVE, Boston Scientific Corporation) with current state-of-the-art single-use and reusable flexible ureteroscopes.

Methods:
LithoVue Elite is the latest single-use flexible ureteroscope with improved functionality and a novel ability to measure intrarenal pressure. In this study, we compare LVE with 3 single-use ureteroscopes (Pusen, WiScope, and Anqing) and 1 reusable ureteroscope (Olympus). A bench model was used to measure the physical characteristics (tip deflection and irrigation) and image quality (resolution, color accuracy, field of view, and latency). A human cadaveric model was used to score the maneuverability and ergonomics. This abstract shows the least square means from multiway ANOVAs used for analysis.

Results:
The ureteroscopes showed a tip deflection ranging from 243° to 306° with LVE at 279°. The irrigation rates ranged from 28 to 44 mL/min with LVE at 40 ml/min. LVE had a resolution of 6.34 line pair/mm, while the other scopes ranged from 4.86 to 5.77 line pairs/mm. The color accuracy was quantified with Delta E, an image analysis variable, which ranged from 14 to 42 with LVE at 28. The field of view ranged from 55° to 89°, and LVE was at 89°. The image latency ranged from 41 to 92 ms with LVE at 58 ms. Finally, the maneuverability scores ranged from 3.8 to 4.1 of 5 with LVE scoring 4.1 while the ergonomics scores ranged from 2.4 to 4.8 with LVE scoring 4.4.

Conclusion:
LithoVue Elite had above average performance compared with single-use and reusable ureteroscopes and upper-range scores for resolution, field of view, maneuverability, and ergonomics. Accordingly, LVE shows promise; however, further research is needed to assess the impact of intrarenal pressure on the standard of care.