Implementation of a software for REmote COMparison of PARticlE and photon treatment plans: ReCompare
Research output: Contribution to journal › Research article › Contributed › peer-review
Contributors
Abstract
PURPOSE: To guarantee equal access to optimal radiotherapy, a concept of patient assignment to photon or particle radiotherapy using remote treatment plan exchange and comparison - ReCompare - was proposed. We demonstrate the implementation of this concept and present its clinical applicability.
MATERIALS AND METHODS: The ReCompare concept was implemented using a client-server based software solution. A clinical workflow for the remote treatment plan exchange and comparison was defined. The steps required by the user and performed by the software for a complete plan transfer were described and an additional module for dose-response modeling was added.
RESULTS: The ReCompare software was successfully tested in cooperation with three external partner clinics and worked meeting all required specifications. It was compatible with several standard treatment planning systems, ensured patient data protection, and integrated in the clinical workflow.
CONCLUSION: The ReCompare software can be applied to support non-particle radiotherapy institutions with the patient-specific treatment decision on the optimal irradiation modality by remote treatment plan exchange and comparison.
Details
Original language | English |
---|---|
Pages (from-to) | 287-294 |
Number of pages | 8 |
Journal | Zeitschrift fur Medizinische Physik |
Volume | 25 |
Issue number | 3 |
Publication status | Published - Sept 2015 |
Peer-reviewed | Yes |
External IDs
researchoutputwizard | legacy.publication#67980 |
---|---|
PubMed | 25724357 |
ORCID | /0000-0002-7017-3738/work/142253900 |
Scopus | 84941022006 |
Keywords
Keywords
- Algorithms, Internet, Photons/therapeutic use, Radiometry/methods, Radiotherapy Dosage, Reproducibility of Results, Sensitivity and Specificity, Software, Software Validation