Comparison of Automatic Segmentation Algorithms for the Subthalamic Nucleus

Research output: Contribution to journalResearch articleContributedpeer-review

Contributors

  • Witold H. Polanski - , Department of Neurosurgery, TUD Dresden University of Technology (Author)
  • Amir Zolal - , TUD Dresden University of Technology, SRH Wald-Klinikum Gera (Author)
  • Kerim Hakan Sitoci-Ficici - , TUD Dresden University of Technology (Author)
  • Patrick Hiepe - , BrainLAB AG (Author)
  • Gabriele Schackert - , TUD Dresden University of Technology (Author)
  • Stephan B. Sobottka - , Department of Neurosurgery, TUD Dresden University of Technology (Author)

Abstract

Introduction: Various automatic segmentation algorithms for the subthalamic nucleus (STN) have been published recently. However, most of the available software tools are not approved for clinical use. Objective: The aim of this study is to evaluate a clinically available automatic segmentation tool of the navigation planning software Brainlab Elements (BL-E) by comparing the output to manual segmentation and a nonclinically approved research method using the DISTAL atlas (DA) and the Horn electrophysiological atlas (HEA). Methods: Preoperative MRI data of 30 patients with idiopathic Parkinson's disease were used, resulting in 60 STN segmentations. The segmentations were created manually by two clinical experts. Automatic segmentations of the STN were obtained from BL-E and Advanced Normalization Tools using DA and HEA. Differences between manual and automatic segmentations were quantified by Dice and Jaccard coefficient, target overlap, and false negative/positive value (FNV/FPV) measurements. Statistical differences between similarity measures were assessed using the Wilcoxon signed-rank test with continuity correction, and comparison with interrater results was performed using the Mann-Whitney U test. Results: For manual segmentation, the mean size of the segmented STN was 133 ± 24 mm 3 . The mean size of the STN was 121 ± 18 mm 3 for BL-E, 162 ± 21 mm 3 for DA, and 130 ± 17 mm 3 for HEA. The Dice coefficient for the interrater comparison was 0.63 and 0.54 ± 0.12, 0.59 ± 0.13, and 0.52 ± 0.14 for BL-E, DA, and HEA, respectively. Significant differences between similarity measures were found for Dice and Jaccard coefficient, target overlap and FNV between BL-E and DA; and FPV between BL-E and HEA. However, none of the differences were significant compared to interrater variability. The analysis of the center of gravity of the segmentations revealed that the BL-E STN ROI was located more medially, superior and posterior compared to other segmentations. Regarding the target overlap for beta power within the STN ROI included with the HEA, the BL-E segmentation showed a significantly higher value compared to manual segmentation. Conclusion: Automatic image segmentation by means of the clinically approved software BL-E provides STN segmentations with similar accuracy like research tools, and differences are in the range of observed interrater variability. Further studies are required to investigate the clinical validity, for example, by comparing segmentation results of BL-E with electrophysiological data.

Details

Original languageEnglish
Pages (from-to)256-262
Number of pages7
JournalStereotactic and functional neurosurgery
Volume98
Issue number4
Publication statusPublished - 1 Aug 2020
Peer-reviewedYes

External IDs

PubMed 32369819
ORCID /0000-0002-6603-5375/work/148606650

Keywords

ASJC Scopus subject areas

Keywords

  • Automatic image segmentation, Brain atlas, Deep brain stimulation, Neuromodulation, Subthalamic nucleus