Betrieb eines multidiziplinar genutzten, interventionellen 0,5-Tesla- Magnetresonanztomographen: Technische und logistische Probleme sowie erste klinische Ergebnisse

Research output: Contribution to journalResearch articleContributedpeer-review

Contributors

  • F. Schmidt - , Leipzig University (Author)
  • J. Dietrich - , Leipzig University (Author)
  • J. P. Schneider - , Leipzig University (Author)
  • J. Thiele - , Leipzig University (Author)
  • S. Lieberenz - , Leipzig University (Author)
  • A. Werner - , Leipzig University (Author)
  • W. Gründer - , Leipzig University (Author)

Abstract

Multidisciplinary usage of a MRI system with a superconducting 0.5-T magnet (Signa SP, General Electric) with a vertical gap suited for diagnostic and therapeutic interventions raises complex problems. The MR equipment, including a special localizing system and the instruments for diagnostic and therapeutic interventions, is described. Before putting the system into operation tests were necessary to check MR compatibility of various materials and instruments and to build some auxiliary equipment. We report on the coordinating activities of the radiologist in the context of MR use by different medical specialities. Within the course of 12 months, 428 examinations/interventions of different kinds were carried out, among them 75 functional examinations of the spine or of joints, 31 diagnostic biopsies, 23 cerebral biopsies, and 23 operations. The special design of the Signa SP allows the whole scale of functional examinations, up to complete neurosurgical interventions, in the sitting position and under nearly real- time imaging control.

Translated title of the contribution
Technological and logistic problems and first clinical results of an interventional 0.5-T MRI system used by various medical specialities

Details

Original languageGerman
Pages (from-to)173-184
Number of pages12
JournalRadiologe
Volume38
Issue number3
Publication statusPublished - Mar 1998