Algorithm for Rapid Exclusion of Clinically Relevant Plasma Levels of Direct Oral Anticoagulants in Patients Using the DOAC Dipstick: An Expert Consensus Paper
Research output: Contribution to journal › Research article › Contributed › peer-review
Contributors
Abstract
BACKGROUND: With the widespread use of direct oral anticoagulants (DOACs), there is an urgent need for a rapid assay to exclude clinically relevant plasma levels. Accurate and rapid determination of DOAC levels would guide medical decision-making to (1) determine the potential contribution of the DOAC to spontaneous or trauma-induced hemorrhage; (2) identify appropriate candidates for reversal, or (3) optimize the timing of urgent surgery or intervention.
METHODS AND RESULTS: The DOAC Dipstick test uses a disposable strip to identify factor Xa- or thrombin inhibitors in a urine sample. Based on the results of a systematic literature search followed by an analysis of a simple pooling of five retrieved clinical studies, the test strip has a high sensitivity and an acceptably high negative predictive value when compared with levels measured with liquid chromatography tandem mass spectrometry or calibrated chromogenic assays to reliably exclude plasma DOAC concentrations ≥30 ng/mL.
CONCLUSION: Based on these data, a simple algorithm is proposed to enhance medical decision-making in acute care indications useful primarily in hospitals not having readily available quantitative tests and 24/7. This algorithm not only determines DOAC exposure but also differentiates between factor Xa and thrombin inhibitors to better guide clinical management.
Details
Original language | English |
---|---|
Pages (from-to) | 770-777 |
Number of pages | 8 |
Journal | Thrombosis and haemostasis |
Volume | 124 |
Issue number | 8 |
Publication status | Published - Aug 2024 |
Peer-reviewed | Yes |
External IDs
Scopus | 85184774335 |
---|
Keywords
Keywords
- Humans, Factor Xa Inhibitors/blood, Administration, Oral, Algorithms, Consensus, Hemorrhage, Predictive Value of Tests, Drug Monitoring/methods, Anticoagulants/therapeutic use, Clinical Decision-Making, Antithrombins, Reagent Strips, Tandem Mass Spectrometry, Reproducibility of Results