Acquired Thrombocytopenia in Contemporary Transcatheter Aortic Valve Prosthesis

Research output: Contribution to journalResearch articleContributedpeer-review

Contributors

Abstract

BACKGROUND:  Postprocedural thrombocytopenia is a known phenomenon following transcatheter aortic valve implantation (TAVI). The aim of this study is to evaluate whether postinterventional platelet kinetics differ when comparing the current generation of balloon-expandable valve (BEV) and self-expanding valve (SEV) prostheses.

METHODS:  We performed a retrospective analysis of patients undergoing TAVI at our facility between 2017 and 2019. Patients were stratified according to the type of prosthesis used: BEV or SEV. Hematocrit-corrected platelet counts were calculated to account for dilution. Nadir platelet counts (lowest recorded platelet count), drop platelet counts (DPC; highest relative platelet drop from baseline), and severity of thrombocytopenia during the discourse and at discharge were assessed.

RESULTS:  Of the 277 included patients, 212 received SEV and 65 BEV. BEV patients were younger (81.8 ± 4.4 years vs 79.7 ± 6.8 years, p = 0.03). Further demographic characteristics were similar between groups. Implanted SEV were larger (p < 0.001) and had shorter procedural times (p < 0.01). There were no significant differences in postprocedural discourse. Postinterventional platelet drop was more pronounced in BEV patients in several evaluated metrics: mean DPC (24.3 ± 10.9% vs 18.8 ± 14.8%, p < 0.01), patients with DPC > 30% (n = 19, 29.2%, vs n = 33, 15.6%, p = 0.02), and also when comparing platelet kinetics.

CONCLUSION:  Despite improvements in outcome, the current generation of balloon-expandable TAVI prostheses carries a predisposition for postprocedural thrombocytopenia even when the effects of dilution are accounted for.

Details

Original languageEnglish
JournalThe thoracic and cardiovascular surgeon
Publication statusE-pub ahead of print - 18 Dec 2023
Peer-reviewedYes

External IDs

Scopus 85175789679

Keywords