Transverse and oblique motion effects in fourier domain optical coherence tomography
Publikation: Beitrag in Buch/Konferenzbericht/Sammelband/Gutachten › Beitrag in Konferenzband › Beigetragen › Begutachtung
Beitragende
Abstract
In this study we describe the effects on the interference signal caused by transverse and oblique motion in spatially encoded Fourier Domain Optical Coherence Tomography (FD OCT). The artifacts caused by sample motion were analyzed theoretically and experimentally. The presented results supply an understanding of the SNR degradation caused by sample motion and demonstrates the importance of fast data acquisition in medical imaging.
Details
Originalsprache | Englisch |
---|---|
Titel | 4th European Conference of the International Federation for Medical and Biological Engineering - ECIFMBE 2008 |
Seiten | 816-820 |
Seitenumfang | 5 |
Band | 22 |
Publikationsstatus | Veröffentlicht - 2008 |
Peer-Review-Status | Ja |
Publikationsreihe
Reihe | IFMBE Proceedings |
---|---|
Band | 22 |
ISSN | 1680-0737 |
Konferenz
Titel | 4th European Conference of the International Federation for Medical and Biological Engineering, ECIFMBE 2008 |
---|---|
Dauer | 23 - 27 November 2008 |
Stadt | Antwerp |
Land | Belgien |
Externe IDs
Scopus | 70350645326 |
---|---|
ORCID | /0000-0003-0554-2178/work/142249836 |
ORCID | /0000-0003-2292-5533/work/142256575 |
Schlagworte
ASJC Scopus Sachgebiete
Schlagwörter
- Fourier domain optical coherence tomography, Sample motion, Signal decrease