The microscopy cell (MicCell), a versatile modular flowthrough system for cell biology, biomaterial research, and nanotechnology

Publikation: Beitrag in FachzeitschriftForschungsartikelBeigetragenBegutachtung

Beitragende

  • F. U. Gast - , GeSiM – Gesellschaft für Silizium-Mikrosysteme mbH (Autor:in)
  • P. S. Dittrich - , Leibniz-Institut für Analytische Wissenschaften -ISAS- e.V. (Autor:in)
  • P. Schwille - , Professur für Biophysik, Biotechnologisches Zentrum (BIOTEC) (Autor:in)
  • M. Weigel - , Max Bergmann Zentrum für Biomaterialien Dresden (MBZ) (Autor:in)
  • M. Mertig - , Max Bergmann Zentrum für Biomaterialien Dresden (MBZ) (Autor:in)
  • J. Opitz - , Max Bergmann Zentrum für Biomaterialien Dresden (MBZ) (Autor:in)
  • U. Queitsch - , Max Planck Institute of Molecular Cell Biology and Genetics (Autor:in)
  • S. Diez - , Max Planck Institute of Molecular Cell Biology and Genetics (Autor:in)
  • B. Lincoln - , Universität Leipzig (Autor:in)
  • F. Wottawah - , Universität Leipzig (Autor:in)
  • S. Schinkinger - , Universität Leipzig (Autor:in)
  • J. Guck - , Universität Leipzig (Autor:in)
  • J. Käs - , Universität Leipzig (Autor:in)
  • J. Smolinski - , Leibniz Institute of Polymer Research Dresden (Autor:in)
  • K. Salchert - , Leibniz Institute of Polymer Research Dresden (Autor:in)
  • C. Werner - , Leibniz Institute of Polymer Research Dresden (Autor:in)
  • C. Duschl - , Fraunhofer Institute for Biomedical Engineering (Autor:in)
  • M. S. Jäger - , Fraunhofer Institute for Biomedical Engineering (Autor:in)
  • K. Uhlig - , Fraunhofer Institute for Biomedical Engineering (Autor:in)
  • P. Geggier - , Fraunhofer Institute for Biomedical Engineering (Autor:in)
  • S. Howitz - , GeSiM – Gesellschaft für Silizium-Mikrosysteme mbH (Autor:in)

Abstract

We describe a novel microfluidic perfusion system for high-resolution microscopes. Its modular design allows pre-coating of the coverslip surface with reagents, biomolecules, or cells. A poly(dimethylsiloxane) (PDMS) layer is cast in a special molding station, using masters made by photolithography and dry etching of silicon or by photoresist patterning on glass or silicon. This channel system can be reused while the coverslip is exchanged between experiments. As normal fluidic connectors are used, the link to external, computer-programmable syringe pumps is standardized and various fluidic channel networks can be used in the same setup. The system can house hydrogel microvalves and microelectrodes close to the imaging area to control the influx of reaction partners. We present a range of applications, including single-molecule analysis by fluorescence correlation spectroscopy (FCS), manipulation of single molecules for nanostructuring by hydrodynamic flow fields or the action of motor proteins, generation of concentration gradients, trapping and stretching of live cells using optical fibers precisely mounted in the PDMS layer, and the integration of microelectrodes for actuation and sensing.

Details

OriginalspracheEnglisch
Seiten (von - bis)21-36
Seitenumfang16
FachzeitschriftMicrofluidics and nanofluidics
Jahrgang2
Ausgabenummer1
PublikationsstatusVeröffentlicht - Jan. 2006
Peer-Review-StatusJa

Externe IDs

ORCID /0000-0002-0750-8515/work/142235587

Schlagworte

Schlagwörter

  • Hydrogel valve, Lab-on-chip, Microelectrodes, Microscopy, PDMS microchannel