Electroosmotic flow in microchannels with prismatic elements

Publikation: Beitrag in FachzeitschriftForschungsartikelBeigetragenBegutachtung

Beitragende

  • Yandong Hu - , University of Toronto, Leibniz-Institut für Polymerforschung Dresden (Autor:in)
  • Xiangchun Xuan - , University of Toronto (Autor:in)
  • Carsten Werner - , University of Toronto, Leibniz-Institut für Polymerforschung Dresden (Autor:in)
  • Dongqing Li - , Vanderbilt University (Autor:in)

Abstract

Fundamental understanding of liquid flow through microchannels with 3D prismatic elements is important to the design and operation of lab-on-a-chip devices. In this paper, we studied experimentally and theoretically the electroosmotic flow (EOF) in slit microchannels with rectangular 3D prismatic elements fabricated on the bottom channel wall. The average electroosmotic velocity measured by the current-monitoring technique was found lower than that in a smooth microchannel. This velocity reduction becomes larger in microchannel with larger but less number of the prisms even though the space taken by the prisms are identical. The velocity distribution and streamlines on two typical horizontal planes in the microchannel are measured and visualized by a particle-based technique. These experimental observations are in good agreement with the numerical simulation. The comparison of streamlines near the prisms in the pressure-driven flow with that in the EOF showed that the EOF was more sensitive to the local geometry.

Details

OriginalspracheEnglisch
Seiten (von - bis)151-160
Seitenumfang10
FachzeitschriftMicrofluidics and nanofluidics
Jahrgang3
Ausgabenummer2
PublikationsstatusVeröffentlicht - Apr. 2007
Peer-Review-StatusJa
Extern publiziertJa

Externe IDs

ORCID /0000-0003-0189-3448/work/174430019