Determination of neomycin by LC-tandem mass spectrometry using hydrophilic interaction chromatography

Publikation: Beitrag in FachzeitschriftForschungsartikelBeigetragenBegutachtung

Beitragende

Abstract

A specific, and automated method was developed to quantitate neomycin in human serum. Samples were prepared with an automated solid phase extraction (SPE). The hydrophilic interaction chromatography (HILIC) was used for additional sample cleanup and baseline separation. The analyte neomycin was detected with electrospray ionisation tandem mass spectrometry (ESI-MS-MS). Using a volume of 500 μl biological sample the lower limit of quantification was 100 ng/ml. The described HILIC-MS-MS method is suitable for clinical and pharmcokinetical investigations of neomycin.

Details

OriginalspracheEnglisch
Seiten (von - bis)633-638
Seitenumfang6
Fachzeitschrift Journal of pharmaceutical and biomedical analysis : JPBA
Jahrgang35
Ausgabenummer3
PublikationsstatusVeröffentlicht - 28 Mai 2004
Peer-Review-StatusJa

Externe IDs

Scopus 2342471947
PubMed 15137990
ORCID /0000-0003-1526-997X/work/142247324

Schlagworte

Schlagwörter

  • Aminoglycoside, HILIC, Hydrophilic interaction, LC-MS-MS, Neomycin