Combined modality for ultrasound imaging and electromagnetic tracking
Publikation: Beitrag in Fachzeitschrift › Forschungsartikel › Beigetragen › Begutachtung
Beitragende
Abstract
Computer-assisted interventions (CAI) typically require localization (tracking) of surgical instruments and the patient. For ultrasound (US)-guided interventions, a new compact electromagnetic (EM) field generator enables construction of a combined modality which allows for both, EM tracking and US imaging with one handheld device. In this study, we present a research prototype of such a device and conduct accuracy assessments in a clinical US suite. The results show robust US imaging and EM tracking of the combined device, which emerges as a promising component for US-guided CAI systems.
Details
Originalsprache | Englisch |
---|---|
Fachzeitschrift | Biomedizinische Technik |
Jahrgang | 58 |
Ausgabenummer | SUPPL.1 |
Publikationsstatus | Veröffentlicht - 2013 |
Peer-Review-Status | Ja |
Extern publiziert | Ja |
Schlagworte
ASJC Scopus Sachgebiete
Schlagwörter
- CAI, Compact FG, Computer assisted interventions, Electromagnetic tracking, Ultrasound imaging