Advanced in vitro hemocompatibility assessment of biomaterials using a new flow incubation system

Publikation: Beitrag in FachzeitschriftForschungsartikelBeigetragenBegutachtung

Beitragende

  • Claudia Sperling - , Leibniz Institute of Polymer Research Dresden (Autor:in)
  • Manfred F. Maitz - , Leibniz Institute of Polymer Research Dresden (Autor:in)
  • Vincent Körber - , Leibniz Institute of Polymer Research Dresden (Autor:in)
  • Stefanie Hänsel - , Leibniz Institute of Polymer Research Dresden (Autor:in)
  • Carsten Werner - , Professur für Biofunktionale Polymermaterialien (gB/IPF), Leibniz Institute of Polymer Research Dresden (Autor:in)

Abstract

Physiologically relevant in vitro hemocompatibility assessment of biomaterials remains challenging. We present a new setup that enables standardized whole blood incubation of biomedical materials under flow. A blood volume of 2 mL is recirculated over test surfaces in a custom-made parallel plate incubation system to determine the activation of hemostasis and inflammation. Controlled physiological shear rates between 125 s−1 and 1250 s−1 and minimized contact to air are combined with a natural-like pumping process. A unique feature of this setup allows tracing adhesion of blood cells to test surfaces microscopically in situ. Validation testing was performed in comparison to previously applied whole blood incubation methodologies. Experiments with the newly developed setup showed that even small obstacles to blood flow activate blood (independent of materials-induced blood activation levels); that adhesion of blood cells to biomaterials equilibrates within 5 to 10 min; that high shear rates (1250 compared to 375 s−1) induce platelet activation; and that hemolysis, platelet factor 4 (PF4) release and platelet loss - but not thrombin formation – depend on shear rate (within the range investigated, 125 to 1250 s−1).

Details

OriginalspracheEnglisch
Aufsatznummer213555
FachzeitschriftBiomaterials advances
Jahrgang153
PublikationsstatusVeröffentlicht - Okt. 2023
Peer-Review-StatusJa

Externe IDs

PubMed 37478769
ORCID /0000-0003-0189-3448/work/161890256

Schlagworte

Schlagwörter

  • Hemocompatibility, In vitro testing, Perfusion chamber, Shear stress, Whole blood