Sound localisation of low- and high-frequency sounds in cochlear implant users with single-sided deafness
Publikation: Beitrag in Fachzeitschrift › Forschungsartikel › Beigetragen › Begutachtung
Beitragende
Abstract
OBJECTIVE: Localisation of low- and high-frequency sounds in single-sided deaf cochlear implant users was investigated using noise stimuli designed to mitigate monaural localisation cues.
DESIGN: Within subject design. Sound source localisation was tested in the horizontal plane using an array of seven loudspeakers along the azimuthal angle span from -90° to +90°. Stimuli were broadband noise and high- and low-frequency noise.
STUDY SAMPLE: Twelve adult subjects with single-sided deafness participated in the study. All had normal hearing in the healthy ear and were supplied with a cochlear implant (CI) in their deaf ear.
RESULTS: With broadband noise, the mean angular localisation error was 39° in aided condition as compared to a median angular error of 83.6° when the speech processor was not worn. For high-frequency noise, the median angular error was 30° and for low-frequency noise, it was 46° in the CI-aided condition.
CONCLUSIONS: Single-sided deaf CI users show the best sound localisation for high-frequency sounds. This supports the view that interaural level differences are dominant for sound localisation in these listeners. Nonetheless, a limited ability to localise low-frequency sounds was observed, which may be based on the supportive perception of interaural time differences.
Details
Originalsprache | Englisch |
---|---|
Seiten (von - bis) | 71-78 |
Seitenumfang | 8 |
Fachzeitschrift | International journal of audiology |
Jahrgang | 62 |
Ausgabenummer | 1 |
Publikationsstatus | Veröffentlicht - Jan. 2023 |
Peer-Review-Status | Ja |
Externe IDs
Scopus | 85124186599 |
---|
Schlagworte
Schlagwörter
- Adult, Cochlear Implantation, Cochlear Implants, Deafness/diagnosis, Humans, Sound Localization, Speech Perception