Feedback suppression in a mechanically coupled, short signal path sensor-actuator system for use as a hearing implant

Publikation: Beitrag in Buch/Konferenzbericht/Sammelband/GutachtenBeitrag in KonferenzbandBeigetragenBegutachtung

Beitragende

Abstract

We examine a 10:1 scale model of a recently proposed type of integrated sensor-actuator system to be used as part of a middle-ear hearing implant, and show that this inherently unstable system can be controlled by conventional methods of digital feedback suppression. Sensor and actuator of the model are mechanically coupled by a free-floating hard metal frame about 30 mm in length, corresponding to 3 mm original scale. Due to the direct mechanical coupling, the resulting electroacoustic system exhibits strong feedback and high-frequency oscillation will occur even at small gain. We show that the feedback of this system can be successfully controlled by delaying the signal in the forward path and then applying adaptive and active filtering. Through the use of a least-mean-square adaptive filter combined with active digital filtering, an increase in maximum stable gain (MSG) of up to 43 dB and a functional gain of up to 35 dB (maximum) and 24.7 dB (mean) are achieved.

Details

OriginalspracheEnglisch
Titel 2015 38th International Conference on Telecommunications and Signal Processing (TSP)
Herausgeber (Verlag)Institute of Electrical and Electronics Engineers Inc.
ISBN (elektronisch)978-1-4799-8498-5, 978-1-4799-8497-8
PublikationsstatusVeröffentlicht - 9 Okt. 2015
Peer-Review-StatusJa

Publikationsreihe

ReiheInternational Conference on Telecommunications and Signal Processing (TSP)

Konferenz

Titel2015 38th International Conference on Telecommunications and Signal Processing
KurztitelTSP 2015
Veranstaltungsnummer38
Dauer9 - 11 Juli 2015
OrtClarion Congress Hotel Prague
StadtPrague
LandTschechische Republik

Schlagworte

Schlagwörter

  • Adaptive Filters, Feedback Control, Hearing Implant, Sensor-Actuator Systems