Early outcomes of aortic valve replacement with Perceval PLUS sutureless valve: results of the prospective multicentric MANTRA study
Publikation: Beitrag in Fachzeitschrift › Forschungsartikel › Beigetragen › Begutachtung
Beitragende
- Klinik für Kardiochirurgie (am Herzzentrum)
- Institute for Cardiovascular Diseases Dedinje
- Hospital de Santa Maria Lisbon
- Seoul National University Hospital
- Ospedale del Cuore Di Massa
- Universitätsklinikum Essen
- Universita della Campania Luigi Vanvitelli
- University Hospitals Cleveland Medical Center
- Ochsner Health System
- Paracelsus Medizinische Privatuniversität Nürnberg
- Fondazione Poliambulanza
- Herzzentrum Dresden GmbH – Universitätsklinik
- Klinikum Oldenburg
- Pusan National University Yangsan Hospital
- Policlinico Paolo Giaccone
- Catharina Hospital Eindhoven
- Azienda Ospedaliera - Universitaria Città della Salute e della Scienza di Torino
- East Limburg Hospital
- Universitair Ziekenhuis (UZ) Leuven
Abstract
BACKGROUND: The aim of this study is to report the preliminary real-word clinical and hemodynamic performance from the MANTRA study in patients undergoing aortic valve replacement with Perceval PLUS sutureless valve.
METHODS: MANTRA is an ongoing "umbrella" prospective, multi-center, international post-market study to collect real-life safety and performance data on Corcym devices (Corcym S.r.l, Saluggia, Italy). Clinical and echocardiographic outcomes were collected preoperatively, at discharge and at each follow up. KCCQ-12 and EQ-5D-5L quality of life questionnaires were collected preoperatively and at 30-days.
RESULTS: A total of 328 patients underwent aortic valve replacement with Perceval PLUS in 29 International institutions. Patients were enrolled from July 2021 to October 2023 and enrollment is still ongoing. Mean age was 71.9 ± 6.4 years, mean EuroSCORE II was 2.9 ± 3.9. Minimally invasive approach was performed in 44.2% (145/328) of patients; concomitant procedures were done in 40.8% (134/328) of cases. Thirty-day mortality was 1.8% (6/328) and no re-interventions were reported. Pacemaker implant was required in 4.0% (13/328) of the patients. The assessment of the functional status demonstrated marked and stable improvement in NYHA class in most patients at 30-day follow-up, with significant increase of KCCQ-12 summary score (from 58.8 ± 23.0 to 71.8 ± 22.1, p < 0.0001) and EQ-5D-5L VAS score (from 64.5 ± 20.4 to 72.6 ± 17.5, p < 0.0001). Mean pressure gradient decreased from 46.2 ± 17.3 mmHg to 10.1 ± 4.7 mmHg at 30-day follow-up. Low or no incidence of moderate-to-severe paravalvular or central leak was reported.
CONCLUSIONS: Preliminary results demonstrate good clinical outcomes and significant improvement of Quality of Life at 30-days, excellent early hemodynamic performance within patient implanted with Perceval PLUS.
TRIAL REGISTRATION: The MANTRA study has been registered in ClinicalTrials.gov (NCT05002543, Initial release 26 July 2021).
Details
Originalsprache | Englisch |
---|---|
Aufsatznummer | 340 |
Seiten (von - bis) | 340 |
Fachzeitschrift | Journal of cardiothoracic surgery |
Jahrgang | 19 |
Ausgabenummer | 1 |
Publikationsstatus | Veröffentlicht - 21 Juni 2024 |
Peer-Review-Status | Ja |
Externe IDs
PubMedCentral | PMC11191267 |
---|---|
Scopus | 85196651454 |
Schlagworte
Schlagwörter
- Aged, Aged, 80 and over, Aortic Valve Stenosis/surgery, Aortic Valve/surgery, Female, Heart Valve Prosthesis, Heart Valve Prosthesis Implantation/methods, Hemodynamics/physiology, Humans, Male, Middle Aged, Prospective Studies, Prosthesis Design, Quality of Life, Sutureless Surgical Procedures/methods, Treatment Outcome