Correction of geometrical effects of a knife-edge slit camera for prompt gamma-based range verification in proton therapy

Publikation: Beitrag in FachzeitschriftForschungsartikelBeigetragenBegutachtung

Beitragende

Abstract

Prompt gamma (PG) based range verification can potentially reduce the safety margins in proton therapy. A knife-edge slit camera has been developed in this context using analytical PG simulations as reference for absolute range verification during patient treatment. Geometrical deviations between measurement and simulation could be observed and have to be corrected for in order to improve the range retrieval of the system. A geometrical correction model is derived from Monte Carlo simulations in water. The influence of different parameters is tested and the model is validated in a dedicated benchmark experiment. We found that the geometrical correction improves the agreement between measured and simulated PG profiles resulting in an improved range retrieval and higher accuracy for absolute range verification. An intrinsic offset of 1.4 mm between measurement and simulation is observed in the experimental data and corrected in the PG simulation. In summary, the absolute range verification capabilities of a PG camera have been improved by applying a geometrical correction model.

Details

OriginalspracheEnglisch
Aufsatznummer25
FachzeitschriftInstruments
Jahrgang2
Ausgabenummer4
PublikationsstatusVeröffentlicht - Dez. 2018
Peer-Review-StatusJa

Externe IDs

ORCID /0000-0003-4261-4214/work/147143125

Schlagworte

ASJC Scopus Sachgebiete

Schlagwörter

  • Monte carlo simulation, Prompt gamma, Proton therapy, Range verification