Correcting magnetic resonance k -space data for in-plane motion using an optical position tracking system
Publikation: Beitrag in Fachzeitschrift › Forschungsartikel › Beigetragen › Begutachtung
Beitragende
Abstract
Purpose: Motion is a major confound of image quality in MRI. A method of retrospectively correcting the effects of rotations and translations on the acquired k -space data is presented. Methods: In two phantom experiments of well-controlled translation and rotation, two MRI-compatible infrared cameras recorded motion data that were used subsequently to correct the position and phase of recorded k -space samples. Motion data can be acquired with a temporal resolution of 60 Hz and spatial accuracy of 0.1 mm for translations and 0.2° for rotations. Results: Significant improvements of image quality are demonstrated. Conclusions: The key advantages of the technique are that it is easy to implement, does not interfere with or complicate MR data acquisition, and is capable of correcting distortions within a single slice. Therefore, the technique has the potential to improve upon approaches that rely on the registration or realignment of successive imaging slices.
Details
Originalsprache | Englisch |
---|---|
Seiten (von - bis) | 5580-5585 |
Seitenumfang | 6 |
Fachzeitschrift | Medical physics |
Jahrgang | 36 |
Ausgabenummer | 12 |
Publikationsstatus | Veröffentlicht - 2009 |
Peer-Review-Status | Ja |
Extern publiziert | Ja |
Externe IDs
PubMed | 20095270 |
---|---|
ORCID | /0000-0001-8870-0041/work/142251358 |
Schlagworte
ASJC Scopus Sachgebiete
Schlagwörter
- Image artifacts, Infrared cameras, Motion artifacts, MRI, Retrospective motion correction